Specifications

Connection Vacuum Inlet Flange DN 160 CF (8")
Connection vacuum outlet flange DN 25 ISO-KF (NW25)
Cooling Method Air Cooling (convection) Air Cooling (forced - fan) Water Cooling
Supply Voltage DC 48V
Power Total Rated 600.00
Pumping Speed Air Nitrogen (@50Hz or max @default setting) 0.680
Pressure Ultimate 3.000 × 10⁻⁸

TURBOVAC 850 i - DN 160 CF

Part Number: 840081V1000

Variant

Explore the range of variants given below to compare the product

Short Description Part Number variantTmpPerformanceModel
840071V1000 TURBOVAC i - frontend 1 (basic)
840071V2000 TURBOVAC i - frontend 2 (Anybus/Fieldbus extention)
840071V3000 TURBOVAC iX - Xtention box controller
840071V3300 TURBOVAC iX - Xtention box controller
840072V1000 TURBOVAC i IND - Industrial version - frontend 1 (basic)
840073V1000 TURBOVAC i(X) UHV - Ultra High Vacuum version - frontend 1 (basic)
840073V2000 TURBOVAC i(X) UHV - Ultra High Vacuum version - frontend 2 (Anybus/Fieldbus extention)
840081V1000 TURBOVAC i - frontend 1 (basic)
840081V2000 TURBOVAC i - frontend 2 (Anybus/Fieldbus extention)
840081V3000 TURBOVAC iX - Xtention box controller
840081V3300 TURBOVAC iX - Xtention box controller
840083V1000 TURBOVAC i(X) UHV - Ultra High Vacuum version - frontend 1 (basic)
840083V2000 TURBOVAC i(X) UHV - Ultra High Vacuum version - frontend 2 (Anybus/Fieldbus extention)

TURBOVAC i/iX

TURBOVAC i is a well-proven turbomolecular pump platform ranging from 90 to 1450 l/s. TURBOVAC i comes with the latest hybrid bearing design offering a permanent magnet upper bearing paired with an oil-free lower mechanical bearing offering the cleanest high vacuum conditions.

 

With TURBOVAC i we offer variants covering UHV (Ultra High Vacuum), throughput IND (Industrial) as well as iR (radiation robust) variants with remote electronics. TURBOVAC i pumps are available with a compact interface offering the control of up to two accessories, basic RS485, USB and DIGITAL I/O control. The TURBOVAC i series also offers the full spectrum of field bus support. The universal connection called "Anybus" enables communication via Profibus, Ethernet IP, ProfiNET as well as EtherCAT through selectable modules.

 

TURBOVAC iX also offers the Anybus interface, the ability to control your fore vacuum pump, and up to 6 accessories, a gauge interface turning the pump into a full vacuum control solution. Along with single inlet pumps, we offer a wealth of multi-inlet, cartridge pumps, and customized products via our Customer Focused Engineering team - please come and talk to our sales teams.

 

TURBOVAC i pumps, in general, come stand-alone - the suitable power supply unit, comms/power cables, and cooling options need to be ordered separately. TURBOVAC 1350, 1450 and the 850, 950 IND variants always come already equipped with a water cooling block.

TURBOVAC 850 i - DN 160 CF

The pump can be operated in any orientation. This product comes without any additional cooling device - an air cooler or a water cooling block need to be ordered separately. As an electronic interface this product comes with the basic compact interface.

Features and Benefits

  • Superior performance - high pumping speed for light gases combined with high compression
  • Stress-free operation
  • High flexibility in our platform for standard and especially in customized products
  • Higher resistance to particles
  • Smart interfaces and intuitive software platform with LeyAssist
  • Operation in any orientation

General Product Information

Accessories / Scope of Supply TURBOVAC i pumps come stand-alone - the suitable power supply unit, comms/power cables, and cooling options need to be ordered separately. IND industrial variants come with an equipped water cooling unit and are water cooling only.
Connection Cooling 6x1 mm flexible hose quick connection
1/8" BSPP (G) Female
Connection vacuum outlet flange DN 25 ISO-KF (NW25)
Connection Vacuum Inlet Flange DN 160 CF (8")
Cooling Method Air Cooling (convection)
Air Cooling (forced - fan)
Water Cooling
Dimension Height 0.292
Dimension Length 0.266
Dimension Width 0.241
IP Rating IP40
Supply Frequency DC Voltage
Supply Voltage DC 48V
Weight 19.60
Dimension Comments The dimensions length, width and height try to describe the outer shape of this particular product in its typical orientation when standing on the ground. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models.
Connection Inlet / Outlet Flange DN 160 CF (8") | DN 25 ISO-KF (NW25)
ProductCertification CE
UL
CSA
connectionPurgeGasBallastVentConnector 1/8" BSPP (or G 1/8") Female
1/8" BSPP (or G 1/8") Female

Environmental Conditions

Temperature Ambient Operation Max 40.00
Temperature Ambient Storage Max 70.00
Temperature Ambient Operation Min 5.00
Temperature Ambient Storage Min -15.00
Magnetic Field Resistance Max (covering worst direction - come and consult us for more details) 9.000 × 10⁻³ T

Classification

Dry or Wet Pump Dry
Product Pressure Range High Vacuum: 1·10⁻³ - 1·10⁻⁷ mbar
Ultra High Vacuum: 1·10⁻⁷ - 1·10⁻¹¹ mbar
High / Medium / Rough Vacuum Classification High Vacuum Pump
Vacuum Pump Technology Turbomolecular Pump - hybrid bearing oil free

Performance Data

Compression Ratio Air Nitrogen 1.000 × 10¹¹
Compression Ratio Helium 1.000 × 10⁸
Compression Ratio Hydrogen 4.000 × 10⁶
Power Total Rated 600.00
Pressure Ultimate 3.000 × 10⁻⁸
Pumping Speed Air Nitrogen (@50Hz or max @default setting) 0.680
Pumping Speed Argon 0.610
Pumping Speed Helium 0.790
Pumping Speed Hydrogen 0.710
Ramp Up Time 150.00
Nominal rotational speed 820.00
Performance Comment The compression of a TMP is a value describing how good the design can compress a gas type at special conditions. Compression data is measured only using the CF flange variants. The values are stated for all family members. For some pump classes we are stating different pumping speeds ISO > CF flange as small differences in geometry influence their performance leading to a difference >5%.
Gas Throughput Air Nitrogen Max 1.40
Gas Throughput Helium Max 2.00
Gas Throughput Max Hydrogen 2.00
Gas Throughput Argon Max 0.350
Pressure Backing Max 1500.00
Compression Ratio Argon 1.000 × 10¹¹

Connectivity - Communication Interface

Communication Interface Protocol USS Universal Serial Interface Protocol
Communication Interfaces Customer RS485 (serial)
REMOTE X1 (Parallel I/O)
USB
Connection electrical comms interface connector DSUB 9-PIN - Male
DSUB HD 15-PIN - Female (high density)
USB type B - Female

Variants

Installation Any orientation
variantTmpPerformanceModel TURBOVAC i - frontend 1 (basic)

Operation Conditions - Limitations

Flow Cooling Water Min 1.667 × 10⁻⁵
Temperature Cooling Water Max 35.00
Temperature Cooling Water Min 15.00
Flow Cooling Water Max 3.000 × 10⁻⁵
Pressure Permissible Cooling Water - Min and Max 4.000 × 10⁵
4.000 × 10⁵
Product Title Part Number Price Add to cart
Cable Accessory - TURBOVAC i / TURBOLAB M8-M8 2m 800110V0016 -
Communication cable - USB cable 2.0 type A/B 1.8m 800110V0108 -
Vent valve - 24V DC - 1/8" BSPP (G) Male (450mm) 800120V0012 -
Start-stop switch for TURBOVAC i/iX 800110V0021 -
TURBO.CONTROL i 800100V0004 -
Water Cooling Unit - TURBOVACi 1/4" BSPP (G) -- tube - (850/950/1350/1450) 800135V0007 -
Cable Power DC - Extension cable TURBOVACi 850-1450 3m 48V DC (TP800 only) 800096V0303 -
Cable Power DC - Extension cable TURBOVACi 850-1450 5m 48V DC (TP800 only) 800096V0503 -
Mounting Kit TP800 TURBOVACi 850-1450 (washers and bolts enclosed) 800100V0110 -
Connector Accessory Y-connector - TURBOVAC i / TURBOLAB M8 800110V0020 -
Air Cooling Unit - TURBOVAC 850 - 950i radial 800136V0011 -
Cable Power DC - TURBOVACi 850-1450 Supply 48V DC open end 1m (external power supply) 800096V0101 -
Cable Power DC - TURBOVACi 850-1450 Supply 48V DC open end 3m (external power supply) 800096V0301 -
Inlet screen coarse - DN 160 CF 800132V0031 -
Mounting kit TURBOVAC - DN 160 CF 800134V0031 -
Power Failure Vent Valve - 24V DC - 1/8" BSPP (G) Male (450mm) 800120V0022 -
Purge Valve - EM 24V DC - 1/8" BSPP (G) Male (450mm) 800120V0013 -
Splinter guard - DN 160 CF fine 800132V0032 -
Cable Power DC - TURBOVACi 850 - 1450 Supply 48V DC open end 5m (external power supply) 800096V0501 -
Air Gas Filter - TURBOVAC 1/8" BSPP (G) Male 800110V0022 -
Cable Accessory - TURBOVAC i / TURBOLAB M8-M8 5m 800103V0003 -
THERMOVAC TTR 91 RN - DN 16 ISO-KF 21L1011100 -
PENNINGVAC PTR 90 RN - DN 40 CF 23L0041100 -